It was reported that during a da vinci left hemicolectomy procedure, the sealing function of the endowrist one vessel sealer instrument seemed intermittent or would take an extremely long time for cut indicator to sound.On (b)(4) 2015, intuitive surgical, inc.(isi) obtained following information from isi clinical sales representative (csr): according to the csr, the vessel sealer instrument seemed to have issues from the beginning.The sealing function was intermittent or would take an extremely long time for the two beeps which typically occurs after sealing.The vessel sealer instrument did not appear to be sealing while the generator was beeping for a long time.There was no smoke and no bubbles coming out while it kept beeping.There were no error messages.The sealing cycle was sometimes in excess of 20+ seconds.According to the csr, the surgeon was dissecting omentum.There was little or no tissue effect sometimes.The surgeon utilized the cut function and there was bleeding, which was controlled quickly with either an additional seal or monopolar curved scissors.No blood transfusion was required.A new vessel sealer instrument was used to complete the procedure.There was no patient injury.
|
Intuitive surgical, inc.(isi) has received the instrument involved with this complaint and completed investigations.Failure analysis investigation confirmed the customer reported complaint of intermittent sealing.The instrument was placed on an in-house system and passed self-check.The cut test passed and there was intermittent sealing.The gap gage test failed.This complaint is being reported due to the vessel sealer instrument sealing function being intermittent could likely cause or contribute to an adverse event, if the malfunction were to recur.
|