BECKMAN COULTER COULTER LH 500 HEMATOLOGY ANALYZER; COUNTER, DIFFERENTIAL CELL Back to Search Results
Catalog Number 178833
Device Problem Leak/Splash (1354)
Patient Problem Hemorrhage/Bleeding (1888)
Event Date 05/13/2015
Event Type  malfunction  
Event Description
The customer reported an uncontained diluted blood leak of about 5 ml from the pierce needle on a coulter lh 500 hematology analyzer.The customer was wearing personal protective equipment (ppe) consisting of laboratory coat, glasses and gloves at the time of the event and there was no report of injury or biohazard exposure to open wounds or mucous membranes.Erroneous patient results were not generated and there was no change or effect to patient treatment in connection to the event.There was no impact to patient results and controls.
 
Manufacturer Narrative
A field service engineer (fse) evaluated the instrument on 05/13/2015 and found pinch valve pv21 stuck and not opening preventing the needle bellows from draining.The fse replaced pv21 and the instrument ran without leaks.The repairs were verified per established service procedures.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COULTER LH 500 HEMATOLOGY ANALYZER
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 s kraemer blvd
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th avenue
miami FL 33196
Manufacturer Contact
miranda holland
11800 sw 147th avenue, 32-l23
miami, FL 33196-2031
3053802031
MDR Report Key4812761
MDR Text Key5915757
Report Number1061932-2015-00919
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K042724
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/13/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/02/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number178833
Other Device ID NumberSOFTWARE VERSION: 2A6
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/13/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-