COVIDIEN 36CM PERM CATH KIT X5; DIALYSIS CATHETER Back to Search Results
Model Number 8817748001
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/20/2015
Event Type  malfunction  
Event Description
It was reported to covidien on (b)(6) 2015 that a customer had an issue with a dialysis catheter.The customer stated that there was errhysis (slow bleed) in the joint of the catheter when conducting dialysis, therefore dialysis failed.There was no patient injury.
 
Manufacturer Narrative
Submit date: 06/03/2015 an investigation is currently under way; upon completion the results will be forwarded.Multiple attempts have been made for additional clarification on the incident reported.If additional pertinent information becomes available, the report will be updated.
 
Manufacturer Narrative
The device history record (dhr) was reviewed and no deviations related to this failure mode were found.There were no non-conformances related to the reported issue.No trends or triggers were identified.The product sample consisted of one incomplete kit of product 36cm perm cath kit x5.The incomplete kit came inside a plastic bag and it presented signs of use.The kit was open and contained a catheter and sheath/dilator, oval.Visual inspection was performed and did not present any issue with the components of the kit that was returned.The returned sample was submitted to an underwater test.Both lumens (arterial and venous) extensions did not show bubbles during the test.Manufacturing performs 100% visual inspection per procedure and 100% leak testing per procedure; therefore a leakage would be detected during these processes.The device was returned and evaluated and no issues were found during visual inspection and functional testing.As per the instructions for use, it is necessary to perform a visual inspection before using the device.Do not use the catheter if it appears damaged or defective.The catheter tubing can tear when subjected to excessive force or rough edges.Inspect the catheter frequently for nicks scrapes, and cuts which could impair its performance.Do not use acetone on any part of the catheter.A leak could be caused due to excessive force, sharp objects or due to an inappropriate use of cleaning agents.It must be noted that in-process controls (such as personnel training, incoming quality acceptance testing for raw material, 100% in process visual inspection and visual acceptance sampling) are in place to prevent nonconforming product from leaving the manufacturing operations.This complaint will be used for tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
36CM PERM CATH KIT X5
Type of Device
DIALYSIS CATHETER
Manufacturer (Section D)
COVIDIEN
covidien manufacturing solulfons sa
edificio 820 calle #2 zona franca coyol
alajuela
CS 
Manufacturer (Section G)
COVIDIEN
covidien manufacturing solulfons sa
edificio 820 calle #2 zona franca coyol
alajuela 20101
CS   20101
Manufacturer Contact
thom mcnamara
15 hampshire street
mansfield, MA 02048
5084524811
MDR Report Key4814967
MDR Text Key5833478
Report Number3009211636-2015-00253
Device Sequence Number1
Product Code MSD
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,foreign,user facility
Reporter Occupation Other
Type of Report Initial
Report Date 05/20/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/03/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8817748001
Device Catalogue Number8817748001
Device Lot Number208214X
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received10/12/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-