Catalog Number 8065990601
Device Problem Energy Spectrum Incorrect (1210)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/08/2015
Event Type  malfunction  
Event Description
A technician reported that at the beginning of lasik treatment in the left eye, the energy of the laser was too low, and the treatment was interrupted at 25% of completion.The system was restarted and an energy check and fluence test was conducted, which fixed the issue.Then, the patient was brought back to complete the remaining 75% of the treatment.The treatment was completed with no further issues.At the one day postoperative visit, the patient presented with good near vision, as planned.The patient did not report any symptoms or concerns.No further information is expected.
 
Manufacturer Narrative
Evaluation summary: investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
Evaluation summary: no sample is expected for evaluation.The device history records (dhr) for the device was reviewed.The associated device was released based on the manufacturer's acceptance criteria.Customer reported system displayed a secondary energy too low at the beginning of a case.A review of the technical service onsite showed no abnormalities that could have contributed to this event: laser was successfully verified prior and after the day of the event.The customer decided to complete the surgery after a successful restart of the system, performing an energy check and fluence test.The root cause could not be identified by the investigation.After a restart, the device worked as intended.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLEGRETTO WAVE EYE-Q EXCIMER LASER
Type of Device
OPHTHALMIC EXCIMER LASER SYSTEM
Manufacturer (Section D)
WAVELIGHT GMBH
am wolfsmantel 5
erlangen 91058
GM  91058
Manufacturer (Section G)
WAVELIGHT GMBH
am wolfsmantel 5
erlangen 91058
GM   91058
Manufacturer Contact
eddie darton, md, jd
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8175686660
MDR Report Key4816003
MDR Text Key15128669
Report Number3003288808-2015-05581
Device Sequence Number1
Product Code LZS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P020050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 08/31/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/03/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065990601
Other Device ID NumberV4-1.02
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/06/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age46 YR
-
-