• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE HIGH FLOW HEATED INSUFFLATOR TUBING (5BX); INSUFFLATOR, LAPAROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE HIGH FLOW HEATED INSUFFLATOR TUBING (5BX); INSUFFLATOR, LAPAROSCOPIC Back to Search Results
Catalog Number 0620030407
Device Problems Device Contamination with Chemical or Other Material (2944); Packaging Problem (3007)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/08/2015
Event Type  malfunction  
Event Description
It was reported that there was a piece of metal in the packaging.
 
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Manufacturer Narrative
The foreign material failure mode was confirmed on the unit returned.A small strand of wire was found inside the package.As per risk document most probable root causes are: 1.Manufacturing error; 2.Damage during shipment (inadequate packaging); 3.Damage during shipment (abuse during transit).The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.
 
Event Description
It was reported that there was a piece of metal in the packaging.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW HEATED INSUFFLATOR TUBING (5BX)
Type of Device
INSUFFLATOR, LAPAROSCOPIC
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
kimberly lynch
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key4816627
MDR Text Key19500684
Report Number0002936485-2015-00451
Device Sequence Number1
Product Code HIF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K003792
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,other
Reporter Occupation Other
Type of Report Initial
Report Date 05/08/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/03/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number0620030407
Device Lot Number15044FE2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/15/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received05/08/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-