Catalog Number 723585
Device Problem Device Emits Odor (1425)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/14/2015
Event Type  malfunction  
Event Description
The customer reported a burning smell from the wall cable of the coulter lh 780 hematology analyzer.There was no smoke, sparks, arcing or flames reported by the customer.The fire department was not called and there was no death or injury as a result.There was no medical attention required by any operator.Erroneous patient results were not generated and there was no change or affect to patient treatment in connection with this event.
 
Manufacturer Narrative
A field service engineer (fse) evaluated the instrument on 05/14/2015.The fse found that the power cord on the uninterruptible power supply (ups) was burnt.The fse replaced the cord and the ups, which resolved the issue.The repairs were verified per established procedures.(b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COULTER LH 780 HEMATOLOGY ANALYZER
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 south kraemer blvd.
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th ave
miami FL 33196
Manufacturer Contact
miranda holland
11800 sw 147th ave
32-l23
miami, FL 33196
3053802031
MDR Report Key4823287
MDR Text Key20151873
Report Number1061932-2015-00936
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K061616
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/14/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/07/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number723585
Other Device ID NumberSOFTWARE VER 2D3
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/14/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-