• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECKMAN COULTER COULTER LH 780 HEMATOLOGY ANALYZER; COUNTER, DIFFERENTIAL CELL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECKMAN COULTER COULTER LH 780 HEMATOLOGY ANALYZER; COUNTER, DIFFERENTIAL CELL Back to Search Results
Catalog Number 723585
Device Problem Complete Blockage (1094)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/16/2015
Event Type  malfunction  
Event Description
While a field service engineer (fse) was troubleshooting an unrelated event, the fse discovered a partial clog in the bottom of the diff mixing chamber on a coulter lh 780 hematology analyzer.Erroneous patient results were not generated and there was no change or effect to patient treatment in connection to the event.
 
Manufacturer Narrative
The fse replaced the diff mixing chamber, resolving the issue.The repairs were verified per established service procedures.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COULTER LH 780 HEMATOLOGY ANALYZER
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 s kraemer blvd
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th avenue
miami FL 33196
Manufacturer Contact
miranda holland
11800 sw 147th avenue, 32-l23
miami, FL 33196-2031
MDR Report Key4824172
MDR Text Key5841155
Report Number1061932-2015-00946
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K061616
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/16/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/08/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number723585
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/16/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-