The customer reported that they received questionable results for a total of ten patient samples tested for tina-quant soluble transferrin receptor (stfr).Of the ten samples, nine were found to have erroneous results that were reported outside of the laboratory.The customer initially received a data alarm with one patient sample result.The reaction curve for the result was noted to be "jumpy" and not normal.All other results from that same day were then checked and 9 additional samples were noted to have "jumpy" reaction curves.These 9 sample results were not accompanied by alarms.The samples were repeated and results were different.The first results were reported outside of the laboratory.After noticing the issue, the laboratory communicated that there were faulty results and the repeat results were then reported outside of the laboratory.The reaction curves of other analytes appeared to be ok and these were not retested.Control results were stated to be ok on the morning of (b)(6) 2015.After the first patient results were generated, one control level was analyzed and found to be out of range.All controls were then analyzed approximately 3 hours later and all were within range.The patients were not adversely affected.The stfr reagent lot number was 606449.The reagent expiration date was asked for, but not provided.It was determined that the last calibration of the assay was performed on 05/13/2015.According to product labeling, calibration must be performed every 6 days.
|