MAUDE Adverse Event Report: PLEXUS MANUFACTURING SDN. BHD; PULSE-GENERATOR, PACEMAKER, EXTERNAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PLEXUS MANUFACTURING SDN. BHD; PULSE-GENERATOR, PACEMAKER, EXTERNAL Back to Search Results
Model Number 5392
Device Problems Loose or Intermittent Connection (1371); Device Damaged Prior to Use (2284)
Patient Problem No Patient Involvement (2645)
Event Date 03/11/2015
Event Type  malfunction  
Event Description
It was reported the device was taken out of the box and there was something loose and rattling inside the device.It was noted the device was brand new, never used.The device was returned for repair.There was no patient involvement.
 
Manufacturer Narrative
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.Product event summary: analysis confirmed the reported event; loose screw found stuck to the inside of the test access label when removed for incoming test.Analysis also found one case screw was contaminated and the encoder nuts were not torqued to specification.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 1190 0
MY  11900
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key4829141
MDR Text Key5839787
Report Number3004593495-2015-00094
Device Sequence Number1
Product Code DTE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Biomedical Engineer
Remedial Action Recall
Type of Report Initial
Report Date 03/11/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/09/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/11/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/18/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-1356-2015
Patient Sequence Number1
-
-