Model Number 2090
Device Problems Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Incorrect Measurement (1383); Unable to Obtain Readings (1516)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/17/2015
Event Type  malfunction  
Event Description
It was reported that the programmer was unable to perform threshold testing and that it was freezing during interrogations.The programmer was returned for service.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.(b)(4).
 
Manufacturer Narrative
Product event summary: analysis was unable to confirm the customer comment that the programmer was unable to perform threshold testing and that it was freezing during interrogations, however it was found that the hard drive was at less than 100% and therefore the hard drive was reimaged as a preventive measure.Analysis also noted that the stylus was broken as was the "eject" button on the personal computer memory card international association (pcmcia) card reader and therefore both the stylus and the pcmcia card reader were replaced to resolve.
 
Manufacturer Narrative
Corrected information no eval explain.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM HEART FAILURE
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key4830419
MDR Text Key21064030
Report Number2182208-2015-01774
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative,company representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 04/24/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/10/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number2090
Device Catalogue Number2090
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/13/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/09/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/18/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
2067 RADIOFREQUENCY PROGRAMMER HEAD
-
-