The esu was thoroughly inspected/tested.A technical safety check was performed on the generator.This included an electrical safety check, a function check of each of the equipment's features, and a power output check.All features were/are functioning properly within specifications for the device.No anomalies were found in the review of the unit's device history record (dhr).In conclusion, no equipment problem was found that would have caused or contributed to the event.Most likely there were many factors involved with the pt incident such as the procedure being long, elderly/obese pt's being more susceptible to pressure to the skin, possible reaction to disinfectant used at the site, etc.However, no conclusive determination could be made as to the cause of the event.No trends have been identified with this incident.Erbe usa, inc is now closing the file on this event.
|