Catalog Number LF4200
Device Problems Failure to Cut (2587); Material Protrusion/Extrusion (2979)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/05/2015
Event Type  malfunction  
Event Description
The customer reported that the device failed to cut during the procedure.It was not used on the patient.The device was returned for evaluation and initial inspection discovered that the knife blade was protruding from the jaws.
 
Manufacturer Narrative
(b)(4).One device was received for evaluation.Visual inspection found the knife was protruding from the jaws.The jaws would not open or close due to the protruding knife.The customer's reported condition was confirmed and attributed to user error.Knife trap happens when the blade is extended and the jaws are not completely closed.This allows the knife track to open too wide and the blade moves out of its track.As a safety measure, the knife must be retracted in order to open the jaws.The tissue in the webbing may have prevented the knife from retracting far enough to allow the jaws to open.More frequent cleaning could have reduced the difficulty in activating the knife.Per the ifu, the jaws should be closed and locked before activating the cutter.A device history review was completed and no entries pertinent to the customer's report were noted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGASURE IMPACT
Type of Device
LIGASURE VESSEL SEALING SYSTEM
Manufacturer (Section D)
COVIDIEN LP
5920 longbow drive
boulder
Manufacturer Contact
sharon murphy
5290 longbow dr.
boulder, CO 80301
2034925267
MDR Report Key4835123
MDR Text Key5945576
Report Number1717344-2015-00414
Device Sequence Number1
Product Code GEI
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 06/05/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/10/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2017
Device Catalogue NumberLF4200
Device Lot Number240346LX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/05/2015
Date Manufacturer Received06/05/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/25/2012
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-