It was reported that the physician¿s (b)(4) x50 handheld is not turning on even with a full charge.It was stated that the lock button is off.The handheld and software was received for analysis on 05/20/2015.Product analysis for the returned handheld was completed and approved on 06/09/2015.An analysis was performed on the returned handheld showed no anomalies associated with the main battery identified during the analysis.During the analysis it was identified that the handheld was unable to advance past the welcome screen.The cause for the display anomaly is associated with resistance values being higher than expected in the touch screen circuitry.Once the display was replaced with a known good display, no further anomalies associated with the handheld performance were identified during the analysis.Product analysis for the returned software was completed and approved on 06/09/2015.An analysis was performed on the returned flashcard and no anomalies associated with flashcard software or databases were identified during the flashcard analysis.The flashcard and software performed according to functional specifications.
|