CYBERONICS INC PROGRAMMING SOFTWARE; PROGRAMMING COMPUTER Back to Search Results
Model Number MODEL 250
Device Problems Image Display Error/Artifact (1304); Failure to Power Up (1476); High Test Results (2457)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/18/2015
Event Type  malfunction  
Event Description
It was reported that the physician¿s (b)(4) x50 handheld is not turning on even with a full charge.It was stated that the lock button is off.The handheld and software was received for analysis on 05/20/2015.Product analysis for the returned handheld was completed and approved on 06/09/2015.An analysis was performed on the returned handheld showed no anomalies associated with the main battery identified during the analysis.During the analysis it was identified that the handheld was unable to advance past the welcome screen.The cause for the display anomaly is associated with resistance values being higher than expected in the touch screen circuitry.Once the display was replaced with a known good display, no further anomalies associated with the handheld performance were identified during the analysis.Product analysis for the returned software was completed and approved on 06/09/2015.An analysis was performed on the returned flashcard and no anomalies associated with flashcard software or databases were identified during the flashcard analysis.The flashcard and software performed according to functional specifications.
 
Manufacturer Narrative
Device failure occurred, but did not cause or contribute to a death or serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING COMPUTER
Manufacturer (Section D)
CYBERONICS INC
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS INC
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key4835226
MDR Text Key17309274
Report Number1644487-2015-04869
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Health Professional
Type of Report Initial
Report Date 05/18/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/10/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Device Lot Number1073167
Other Device ID NumberVERSION 8.1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/20/2015
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received05/18/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/22/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-