We received one 746f8 catheter with an attached monoject 3ml syringe with 1.5 ml limited volume for examination.No pressure monitoring device was returned with the product.No fault messages showed up on the laboratory vigilance 2 monitor when the catheter was connected.The thermistor was found to read 37.0 c when submerged into a 37.0 c water bath.The thermistor temperature reading accuracy is +/- 0.3c per the vigilance manual.The catheter ran cco in 37.0 c water bath on the vigilance ii monitor for 5 minutes with no error.The thermistor and thermal filament circuits were continuous, there were no open or intermittent conditions.No visible inconsistencies were observed on the eeprom data.Both the thermistor and thermal filament connectors were opened and found no visible inconsistencies.Resistance value of thermal filament circuit was measured and found to be within specification.All through lumens were patent without any leakage or occlusion.The balloon inflated clear, concentric and remained inflated for more than 5 minutes without leakage.In addition, there was no visible damage observed from the catheter body or returned syringe.A review of the manufacturing records indicated that the product met specifications upon release.The reported event was not confirmed.Although the cause of the complaint could not be determined, there was no indication of a manufacturing defect noted during the analysis.No actions will be taken at this time.
|