Catalog Number VLP12-02B
Device Problem Material Fragmentation (1261)
Patient Problem No Patient Involvement (2645)
Event Date 05/14/2015
Event Type  malfunction  
Event Description
The customer, a biomedical engineer, contacted physio-control to report that a pin had broken off of a therapy cable assembly become lodged in their device's therapy connector assembly.As a result, the device would not detect that the therapy cable was connected and defibrillation was not possible.There was no patient use associated with the reported event.
 
Manufacturer Narrative
(b)(4).Physio-control provided the customer, a biomedical engineer, with technical assistance.It was later confirmed by the biomedical engineer that the therapy connector assembly was replaced to resolve the reported issue.Additionally, the damaged therapy cable assembly was disposed of and a replacement cable was obtained.After observing proper device operation through functional and performance testing the unit was placed back into service for use.Neither the device nor the damaged therapy cable assembly have been returned to physio-control for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 12 DEFIBRILLATOR/MONITOR SERIES
Type of Device
DEFIBRILLATORS, AUTOMATIC, EXTERNAL
Manufacturer (Section D)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer (Section G)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer Contact
jason march
4258674000
MDR Report Key4841916
MDR Text Key15310899
Report Number3015876-2015-00656
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102972
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 05/18/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/12/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberVLP12-02B
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device Age10 YR
Event Location Hospital
Date Manufacturer Received05/18/2015
Date Device Manufactured04/06/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-