Per the sales representative who interviewed the supervising physician following the case: the rca had a 99% ostial blockage.The doctor used a 2.75 mm x 16 mm stent and post dilatation balloon.A 3.0 mm x 8 mm flash was subsequently used.The distal balloon was inflated without issues.When the proximal balloon was inflated, it expanded distal to the stent, damaging the vessel.At least one covered stent was placed to address the damage.The pt was stable following placement of the covered stent.
|
The device used in the procedure was returned to ostial corp for eval.A two page summary of the eval has been attached.No issues were found with the performance of the returned device, though the distal portion of the proximal balloon had clearly been deformed by overexpansion during use.Per the results of the eval, it was found that the most likely root cause of the incident was grossly incorrect positioning of the device prior to inflation.In addition to the case details, ostial corp review the mfg documentation associated with the device lot used during the procedure.No issues were noted that would have contributed to the reported incident.The instructions for use for the prod were also reviewed and it was verified that vessel perforation is listed as a potential complication associated with the procedure and that the user is clearly instructed on how the device should be positioned during use.See scanned pages.
|