Catalog Number 037-00J
Device Problems Device Damaged Prior to Use (2284); Defective Component (2292)
Patient Problems No Consequences Or Impact To Patient (2199); No Patient Involvement (2645)
Event Date 05/25/2015
Event Type  malfunction  
Event Description
The customer alleges that the connector of the bottle was damaged.No patient injury reported.
 
Event Description
The customer alleges that the connector of the bottle was damaged.No patient injury reported.
 
Manufacturer Narrative
(b)(4).A visual, functional and dimensional inspection of the product involved in the complaint could not be conducted since the product was not returned.A device history record reviewed was conducted and showed no issues that may have contributed to any quality issues reported.All process parameters were within specification.All in-process qa inspections shows one non-conformance that has no impact on the quality issue reported.Non-conforming material was sorted for a defect, re-inspected, and found to be acceptable per internal specification.No sample available from the customer to investigate.Teleflex will continue to monitor feedback from the customers on issues related to a damaged bottle connector found at inspection or prior to use.Root cause is unknown.
 
Manufacturer Narrative
Qn#(b)(4).The sample was returned for evaluation.A visual exam was performed and it was found that the bottle was empty due to a drain hole being created by the customer.It was also noticed that the top of the port was slightly bent.No other issues were found.Based on the investigation performed, the reported complaint was confirmed.It was determined that the water bottle was incorrectly packaged when the plastic was still too warm, which resulted in a deformed bottle due to a pressure point during packaging, loading, handling or shipping.Manufacturer will continue to monitor feedback from customers on issues related to a damaged bottle connector found at inspection on water bottle products.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON AQUAPAK 700 SW,760 ML,JAPAN
Type of Device
HUMIDIFIER
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
Manufacturer (Section G)
TELEFLEX MEDICAL
900 west university dr.
arlington heights IL 60004
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9194334854
MDR Report Key4844638
MDR Text Key5934227
Report Number1417411-2015-00104
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/02/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date02/28/2019
Device Catalogue Number037-00J
Device Lot Number021150
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/15/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/21/2015
Initial Date FDA Received06/15/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received07/23/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/23/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-