(b)(4).A review of the complaint history, instructions for use and quality control was conducted for the purpose of this investigation.The product was returned in an open and damaged condition.The tip of the catheter split lengthwise and separated approximately 7mm distal to the bond.Quality inspection final inspection for angiographic catheters states: "verify surface of catheter is free of damage and excess bumps or roughness.Wire braided catheter surface must also be free of exposed wires".Verify distal tip/endhole is rounded smooth, not thin, slanted, or out of round.No splits, nicks, damage, excess material or debris present".This product is shipped with instructions for use which states under precautions: due to thinwall construction, extreme care must be exercised during manipulation and withdrawal.Catheter insertion through a synthetic vascular graft should be avoided whenever possible" / "the possible whiplash effect of the long, soft catheter tip must be considered during selective angiography".The appropriate internal personnel have been notified and we will continue to monitor for similar complaints.Measures have been taken in an effort to reduce the potential for future complaints of this nature.
|