Catalog Number 05.001.202 |
Device Problem
Loss of Power (1475)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Event Description
|
It was reported from (b)(6) that no power was provided to the power module device when in use.It was not reported if the device was used in surgery.There was no delay in the surgical procedure.A spare device was not available for use.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
(b)(6).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
The actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the battery was defective and damaged.Therefore, the reported condition was confirmed.It was further determined that the maximum cycles were completed and that the red led was flashing on the charger.The assignable root cause was determined to be due to wear from normal use and servicing.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|