As they were trying to manipulate the catheter to turn backwards, the catheter tip broke off.They had the tip snared but were having trouble pulling it out.The reporter confirmed the tip was removed from the pt's body.The device will not be returned as the facility is going to keep it for their own investigation.A section of the device did not remain inside the pt's body.According to the initial reporter, the pt did not experience any adverse effects due to this occurrence.
|
(b)(4).A review of the complaint history, instructions for use and qual control was conducted for the purpose of this investigation.No prod was returned to assist with the investigation.Qual control final inspection for angiographic catheters certifies the surface of the catheter is free of damage and excess bumps or roughness.Wire braided catheter surface also verified to be free of exposed wires.The distal tip/endhole is verified for roundness and smoothness, not thin, slanted, or out of round.Also, the device is verified for splits, nicks damage, excess material or dirias present.This prod is shipped with instructions for use which states under precautions: "due to thin wall construction, extreme care must be exercised during manipulation and withdrawal.Catheter insertion through a synthetic vascular graft should be avoided whenever possible."/"the possible whiplash effect of the long, soft catheter tip must be considered during selective angiography." we are inconclusive as to why this failure mode occurred.The appropriate cook inc personnel have been notified and we will continue to monitor for similar complaints.
|