Model Number MX6740 |
Device Problem
Device Slipped (1584)
|
Patient Problem
Hemorrhage/Bleeding (1888)
|
Event Date 01/29/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
Visual inspection of the returned device revealed that there was no saline in the balloon.The balloon was fully inflated with water and pressure was maintained.The inflation indicator performed per specification.No leaks were observed.The inflated balloon diameter was also verified in a go/no go gauge to be within specification.Based on the information provided and the investigation performed, the root cause of the reported event could have been incorrect prep of the balloon.Per the mynx ace instructions for use (ifu), the device needs to be purged of air by drawing vacuum with 2-3 ml of sterile saline prior to use.Failure to purge the device of air during the prep phase can cause the balloon to partially collapse as the air in the system is exposed to additional compressive forces and resulting in the balloon pulling through the arteriotomy.There is no evidence to suggest that the mynx ace vascular closure device did not meet specification or perform as intended per the ifu.The review of the lhr ((b)(4)) indicated that the device lot met all established performance criteria prior to shipment.(b)(4).
|
|
Event Description
|
The following information was reported by the company representative: i was just contacted in regards to this device as the new company representative on the account.It has been sitting on their shelf for return for some time.There is an issue with the balloon and the device pulling out but the doctors do not remember the details.Manual compression was applied for less than 15 minutes.
|
|
Manufacturer Narrative
|
Visual inspection of the returned device revealed that there was no saline in the balloon.The balloon was fully inflated with water and pressure was maintained.The inflation indicator performed per specification.No leaks were observed.The inflated balloon diameter was also verified in a go/no go gauge to be within specification.Based on the information provided and the investigation performed, the root cause of the reported event could have been incorrect prep of the balloon.Per the mynxgrip instructions for use (ifu), the device needs to be purged of air by drawing vacuum with 2-3 ml of sterile saline prior to use.Failure to purge the device of air during the prep phase can cause the balloon to partially collapse as the air in the system is exposed to additional compressive forces and resulting in the balloon pulling through the arteriotomy.There is no evidence to suggest that the mynxgrip device did not meet specification or perform as intended per the instructions for use (ifu).The review of the lhr (f1422003) indicated that the device lot met all established performance criteria prior to shipment.(b)(4).
|
|
Search Alerts/Recalls
|