(b)(4).A device history record (dhr) review did not show issues related to complaint.Complaint cannot be confirmed since the sample was not available for investigation; therefore, it is not possible to determine the root cause for the defect reported and a corrective action for it.However, the manufacturer will continue to monitor and trend relating complaints.
|
(b)(4).One (1) stapler from catalog number 528435 deroyal surgimate 35w (b)(4) case was received used, opened without the original packaging and the lot# was not confirmed.During visual inspection, the components look well assembled and no staple was stuck in the tip of the cartridge.The stapler was received with remaining staples; staples was not observed misaligned and a cracked bottom component.Failure mode jamming was not confirmed with sample received during visual inspection.Despite the condition when the sample was received with a cracked bottom component, a functional inspection was performed with remaining staples (14).There was no quality issues were found during the activation, the device works properly; staples closed & released properly.Therefore, corrective actions are not required at this time.However, the manufacturer will continue to monitor for trends.
|