• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ON-X LIFE TECHNOLOGIES, INC. ON-X MITRAL PROSTHETIC HEART VALVE; MECHANICAL HEART VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ON-X LIFE TECHNOLOGIES, INC. ON-X MITRAL PROSTHETIC HEART VALVE; MECHANICAL HEART VALVE Back to Search Results
Model Number ONXM
Device Problem Interference (1327)
Patient Problem Thrombosis (2100)
Event Date 04/10/2015
Event Type  Injury  
Event Description
Per operative notes: clotted mitral mechanical prosthesis.3rd time re-do mvr.1st mvr in 1987, re-do for clotted valve in 1998, 2nd re-do for clotted valve/pannus from a st jude valve in 2013, then this on-x valve implanted.Valve came out easily with no further dissection suggesting that there may be a chronic infective element in this patient (per surgeon).The latter is not conclusive at this time.Valve thrombosis is by definition in the aats/sts guidelines a valve-related event and is therefore reportable.Valve thrombosis is an expected adverse event, as defined in the objective performance criteria in fda heart valve guidance, and iso 5480, and this occurrence is well-within expected rate.No trend is seen.Correction of the event required re-operation, classified as a "serious injury".Therefore it is reportable as an mdr.
 
Manufacturer Narrative
Device was evaluated by the surgeon who confirmed that the material on the valve blocking the leaflet function was thrombus.In addition, full operative notes were provided stating the surgeon's analysis of the condition.The explanted valve was evaluated by dr.(b)(6) who also provided photos.It was decided that these evaluations were sufficient and the valve did not need to be returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ON-X MITRAL PROSTHETIC HEART VALVE
Type of Device
MECHANICAL HEART VALVE
Manufacturer (Section D)
ON-X LIFE TECHNOLOGIES, INC.
1300 e. anderson ln, bldg b.
austin TX 78752
Manufacturer Contact
walt moeller, director qa/ra
1300 e. anderson ln, bldg b
austin, TX 78752
5123398000
MDR Report Key4861439
MDR Text Key5812259
Report Number1649833-2015-00009
Device Sequence Number1
Product Code LWQ
Combination Product (y/n)N
Reporter Country CodeSF
PMA/PMN Number
P000037
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 06/19/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/19/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date02/23/2017
Device Model NumberONXM
Device Catalogue NumberONXM-25
Other Device ID Number00851788001273
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received05/21/2015
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age34 YR
-
-