(b)(4).Nine (9) cartridges of catalog number 544240 hemolok l clips (b)(4) were received not used, closed in original packaging, lot # 73k1400167 was confirmed with samples received.During visual inspection was observed; all components look well assembled.Functional inspection: the (b)(4) clips were closed (closure test) into the appropriate media tubing p/n 06424-66 according to manufacturing procedures (b)(4); the clips were properly/correctly closed, after the closure test the (b)(4) clips were reviewed/verified (pulled) to duplicated the failure mode.No quality issues were found.Samples received did not confirm the defect reported by the customer "not closing" during visual inspection.In addition, a functional inspection was performed with clips received and the device worked properly.Therefore, a corrective action is not required at this time.However, we will continue to monitor for trends.
|