• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - MALTA ACCESS; STOPCOCK, I.V. SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - MALTA ACCESS; STOPCOCK, I.V. SET Back to Search Results
Catalog Number UMC3325
Device Problems Break (1069); Failure to Disconnect (2541)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Event Description
It was reported that it was difficult to disconnect the perfuser of the stopcock.When the nurse used a clamp kocher tool to disconnect the set, the stopcock broke.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
(b)(4).A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was received for evaluation.Visual inspection was performed and the reported problem was verified.The cause of the reported problem was undetermined.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCESS
Type of Device
STOPCOCK, I.V. SET
Manufacturer (Section D)
BAXTER HEALTHCARE - MALTA
a47 industrial estate, malta b
marsa
MT 
Manufacturer (Section G)
BAXTER HEALTHCARE - MALTA
a47 industrial estate, malta b
marsa
MT  
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key4862526
MDR Text Key15130355
Report Number1416980-2015-26279
Device Sequence Number1
Product Code FMG
Combination Product (y/n)N
Reporter Country CodeFR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Pharmacist
Type of Report Initial
Report Date 06/03/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/23/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/29/2020
Device Catalogue NumberUMC3325
Device Lot Number15C02V913
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/08/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/03/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/04/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-