MAUDE Adverse Event Report: HILL-ROM, INC. VERSACARE BED; A/C POWERED ADJUSTABLE HOSP BED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM, INC. VERSACARE BED; A/C POWERED ADJUSTABLE HOSP BED Back to Search Results
Model Number 3200
Device Problem Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/26/2015
Event Type  malfunction  
Event Description
Hill-rom received a report from a hill-rom tech stating the nurse call was inoperative.The bed was located at the account.There was no pt/user injury reported.This report was filed in our complaint handling system as complaint # (b)(4).
 
Manufacturer Narrative
The hill-rom technician found the nurse call cable damaged.A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this bed from 2012-2015.It is unk if the facility performed any other preventative maintenance on this bed.The tech replaced the nurse call cable to resolve the issue.Based on this info, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSACARE BED
Type of Device
A/C POWERED ADJUSTABLE HOSP BED
Manufacturer (Section D)
HILL-ROM, INC.
batesville IN
Manufacturer Contact
john cummings
1069 state route 46 e
batesville, IN 47006
8129312869
MDR Report Key4862587
MDR Text Key5849679
Report Number1824206-2015-00663
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 05/26/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/18/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number3200
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received05/26/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-