• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR Back to Search Results
Model Number 37601
Device Problems Break (1069); High impedance (1291)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  Injury  
Event Description
It was reported, the patient had open circuits on one side of their system.The reporter thought the open circuits were on the left side.A complete break in a lead was confirmed by x-ray.The brake occurred a couple of centimeters above the extension connection site.The patient was not having any therapy issues with the affected side at the time of this report.The open circuit was discovered during a routine visit.The patient had no falls or trauma and they were getting substantial therapeutic benefit from the unaffected side.A revision was planned for within the next two weeks.No outcome was reported regarding this event.Further follow-up is being conducted to obtain this information.If additional information is received, a follow up report will be sent.
 
Manufacturer Narrative
Product id 37601, serial# (b)(4), implanted: 2015 (b)(6); product type implantable neurostimulator product id 3387s-40, lot# va0tn4t, implanted: 2015 (b)(6); product type lead product id 3708660, serial# (b)(4), implanted: 2015(b)(6); product type extension product id 37642, serial# (b)(4); product type programmer, patient product id 3387s-40, lot# va0tn4t, implanted: 2015 (b)(6); product type lead product id 3708660, serial# (b)(4), implanted: 2015 (b)(6); product type extension.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC NEUROMODULATION
7000 central avenue ne rcw215
minneapolis MN 55432
Manufacturer Contact
diane wolf
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263987
MDR Report Key4866078
MDR Text Key21329562
Report Number3004209178-2015-12138
Device Sequence Number1
Product Code MFR
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H050003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 06/03/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/24/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/14/2016
Device Model Number37601
Device Catalogue Number37601
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/03/2015
Date Device Manufactured12/15/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age00069 YR
-
-