• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS, INC. PROGRAMMING SOFTWARE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS, INC. PROGRAMMING SOFTWARE Back to Search Results
Model Number MODEL 250
Device Problems Failure to Align (2522); Operating System Becomes Nonfunctional (2996)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/02/2015
Event Type  malfunction  
Event Description
It was reported handheld computer was no longer working, and a new programming system was requested by the physician.A replacement system was provided.It was reported that the handheld device was freezing upon interrogation.Mfg report # 1644487-2015-04958 captures a subsequent screen alignment issue.The software has been received by the manufacturer for analysis.However, analysis has not been completed to date.
 
Event Description
Product analysis for the software 250-8.1 was completed and approved on (b)(6) 2015.An analysis was performed on the returned flashcard and showed that no anomalies associated with flashcard software or databases were identified during the flashcard analysis.The flashcard and software performed according to functional specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING SOFTWARE
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key4872678
MDR Text Key6047547
Report Number1644487-2015-05037
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative,company representative
Reporter Occupation Health Professional
Type of Report Initial
Report Date 06/02/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/26/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Device Lot Number1063825
Other Device ID NumberVERSION 8.1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/25/2015
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received07/29/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-