• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY AG ROTAFLOW CENTRIFUGAL PUMP SYSTEM; KFM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY AG ROTAFLOW CENTRIFUGAL PUMP SYSTEM; KFM Back to Search Results
Model Number 20-970
Device Problem Temperature Problem (3022)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/03/2015
Event Type  Injury  
Event Description
It was reported that the rotaflow console emitted a burning smell while replugged in anticipation of a transportation to the scanner.The emergency drive was used before replacing the device.(b)(4).
 
Manufacturer Narrative
Maquet cardiopulmonary (b)(4) provides product failure investigation, analysis and resolution for the device described in this report.The device in question was investigated by a local service technician.The technician noticed a burnt transistor into the alimentation card.Based on the information from service report he replaced the mcp00702711 power supply board and checked the good functioning of ventilator as well as functional control on test circuit.The data is being handled through a designated maquet cardiopulmonary trending and applicable investigation process.Due to this no further investigation initiations will be completed at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAFLOW CENTRIFUGAL PUMP SYSTEM
Type of Device
KFM
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
Manufacturer Contact
michael campbell
kehler strabe 31
rastatt 76437
GM   76437
2229321132
MDR Report Key4884375
MDR Text Key15137154
Report Number8010762-2015-00773
Device Sequence Number1
Product Code KFM
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K991864
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/10/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/30/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20-970
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/10/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age49 YR
-
-