• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CPI - DEL CARIBE INGEVITY; IMPLANTABLE LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CPI - DEL CARIBE INGEVITY; IMPLANTABLE LEAD Back to Search Results
Model Number 7742
Device Problem Material Integrity Problem (2978)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/04/2015
Event Type  malfunction  
Event Description
Boston scientific received information that this right ventricular (rv) lead was an attempted implant.It was reported that the helix mechanism was not functioning as expected.No adverse patient effects were reported.
 
Manufacturer Narrative
(b)(4).Upon receipt at our post market quality assurance laboratory, visual inspection noted the helix was extended and dried blood was present in the helix housing.It was determined that the cathode conductor coil was fractured at the distal end of the terminal pin.Microscopic analysis confirmed that the lead became fractured due to torsional overstress.A laboratory technician tested the helix mechanism and found it was nonfunctional, confirming the clinical observation.Based upon the clinical observations and the laboratory findings, we believe the conductor coil became fractured during attempts to extend/retract the helix.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INGEVITY
Type of Device
IMPLANTABLE LEAD
Manufacturer (Section D)
CPI - DEL CARIBE
guidant puerto rico b. v.
dorado PR
Manufacturer (Section G)
CPI - DEL CARIBE
guidant puerto rico b. v.
dorado PR
Manufacturer Contact
sonali vasekar
4100 hamline ave. n
st. paul, MN 55112
6515824786
MDR Report Key4885024
MDR Text Key6047797
Report Number2124215-2015-08130
Device Sequence Number1
Product Code DTD
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 06/19/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/01/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/07/2017
Device Model Number7742
Other Device ID NumberINGEVITY MRI
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/15/2015
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Manufacturer Received06/19/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/07/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
7736; J277; MISMATCH; 7742
-
-