• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CPI - DEL CARIBE INGEVITY; IMPLANTABLE LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CPI - DEL CARIBE INGEVITY; IMPLANTABLE LEAD Back to Search Results
Model Number 7742
Device Problem Material Integrity Problem (2978)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/02/2015
Event Type  malfunction  
Event Description
Boston scientific received information that during the implant procedure, helix issues were experienced with the right ventricular (rv) lead.As a result, the lead was removed and returned for analysis.No adverse patient effects were reported.
 
Manufacturer Narrative
(b)(4).Upon receipt at our post market quality assurance laboratory, visual inspection noted the helix was retracted and dried blood was present in the helix housing.Resistance testing found the lead was not electrically continuous.Detailed analysis confirmed that the cathode conductor coil was fractured at the distal end of the terminal pin.Microscopic analysis confirmed that the lead became fractured due to torsional overstress.A laboratory technician tested the helix mechanism and found it was nonfunctional, confirming the clinical observation.Based upon the clinical observations and the laboratory findings, we believe the conductor coil became fractured during attempts to extend/retract the helix.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INGEVITY
Type of Device
IMPLANTABLE LEAD
Manufacturer (Section D)
CPI - DEL CARIBE
guidant puerto rico b. v.
dorado PR
Manufacturer (Section G)
CPI - DEL CARIBE
guidant puerto rico b. v.
dorado PR
Manufacturer Contact
sonali vasekar
4100 hamline ave. n
st. paul, MN 55112
6515824786
MDR Report Key4888249
MDR Text Key16561027
Report Number2124215-2015-08546
Device Sequence Number1
Product Code DTD
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 06/29/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/02/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2017
Device Model Number7742
Other Device ID NumberINGEVITY MRI
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/26/2015
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Manufacturer Received06/29/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/31/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
7742; 7741; J066
-
-