Model Number 3700
Device Problem Device Slipped (1584)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/04/2015
Event Type  malfunction  
Event Description
Hill-rom received a report form the account stating the brakes were not holding.The bed was located in labor and delivery at the account.There was no patient/user injury reported.This report was filed in our complain handling system as complaint #(b)(4).
 
Manufacturer Narrative
The technician found the detent assembly was missing some pieces.A search of the hill-rom maintenance showed hill-rom performed preventative maintenance on this bed in 2009-2015.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced the detent assembly to resolve the issue.Based on this information, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFFINITY 4 BED
Type of Device
BIRTHING BED
Manufacturer (Section D)
HILL-ROM, INC.
batesville IN
Manufacturer Contact
tony werner
1069 state route 46 east
batesville, IN 47006
8129312359
MDR Report Key4890621
MDR Text Key21065653
Report Number1824206-2015-00691
Device Sequence Number1
Product Code HDD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K91579
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Unknown
Type of Report Initial
Report Date 06/04/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/29/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number3700
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received06/04/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-