Catalog Number 8065750833
Device Problems Device Emits Odor (1425); Self-Activation or Keying (1557)
Patient Problem No Patient Involvement (2645)
Event Date 06/10/2015
Event Type  malfunction  
Event Description
A clinical manager reported the unit was turned on and immediately turned off on its own.They noticed a burning smell from the lower back of the unit.This was prior to surgery with no patient involvement.
 
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Manufacturer Narrative
The system was examined and the reported event was replicated.The host was replaced were replaced to address the issue.The system was tested and found to meet product specifications.A review of the customer¿s complaint history for the last 24 months did not show any previous complaints of this kind against the system.The system was manufactured on june 25, 2007.Based on qa assessment, the product met specifications at the time of release.The root cause of the reported event can be attributed to a nonconforming host.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITI VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
eddie darton, md, jd
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8175686660
MDR Report Key4891008
MDR Text Key6834179
Report Number2028159-2015-06809
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K120912
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,health professional,oth
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 09/21/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/06/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065750833
Other Device ID Number2.06
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/18/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/25/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-