A computer software error has been verified, on (b)(6) 2015 for the patient¿s model 102 generator from a review of the april programming history database spreadsheet and periodic review algorithm.On (b)(6) 2007 the device was interrogated then programmed, then a system diagnostic was performed.After the system diagnostic the device was interrogated, programmed and a final interrogation the settings were as intended.There is no evidence of a "faulted" diagnostic or communication interruption during this visit.On the next recorded visit (b)(6) 2014 the first interrogation the settings were indicative of a "faulted" diagnostic.There is approximately 7 years of missing programming history and the date of the "faulted" diagnostic is unknown.
|