Model Number MHD6S
Device Problem Particulates (1451)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/10/2015
Event Type  malfunction  
Event Description
It was reported that an unknown particulate was observed inside the sensor before use upon opening the package.There were no patient complications reported.
 
Manufacturer Narrative
Received one flotrac sensor with attached iv and pressure tubing sets.A review of the manufacturing records indicated that the product met specification at the time of release.Unknown black area was found around the fluid path from flotrac sensor housing.The top portion of the flotrac housing was removed for visual examination.The unknown particulate reported was actually the shadow of a void within the uv adhesive bond area (between the sensor housing and the white cable).The void area was outside of the flotrac housing fluid path.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLOTRAC SENSOR
Type of Device
TRANSDUCER, BLOOD-PRESSURE, EXTRAVASCULAR
Manufacturer (Section D)
EDWARDS LIFESCIENCES DR
carretera sanchez, km 18 1/2
parque ind. de itabo
haina. san cristobal
DR 
Manufacturer (Section G)
EDWARDS LIFESCIENCES DR
carretera sanchez, km 18 1/2
parque ind. de itabo
haina. san cristobal
DR  
Manufacturer Contact
lynn selawski
1 edwards way
irvine, CA 92614
9497564386
MDR Report Key4897508
MDR Text Key6833774
Report Number2015691-2015-01617
Device Sequence Number1
Product Code DRS
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K043065
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Literature,Health Professional,User Facility,foreign,health pr
Reporter Occupation Health Professional
Type of Report Initial
Report Date 06/10/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/07/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMHD6S
Device Lot Number59952101
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/26/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/10/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-