MAUDE Adverse Event Report: ATRIUM MEDICAL CORPORATION ATRIUM ICAST COVERED STENT; PROSTHESIS, TRACHEAL, EXPANDABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ATRIUM MEDICAL CORPORATION ATRIUM ICAST COVERED STENT; PROSTHESIS, TRACHEAL, EXPANDABLE Back to Search Results
Model Number 85455
Device Problem Device Slipped (1584)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/02/2015
Event Type  Injury  
Event Description
The surgeon was trying to advance the catheter through a 7 fr cook ansel sheath during a fenestrated graft case.The tip of the catheter had penetrated the introducer sheath but the stent slid off as it was being passed through the hub.
 
Manufacturer Narrative
On completion of the investigation a follow up report will be submitted.Associated file: mdr 1219977-2015-00179.
 
Manufacturer Narrative
Engineering analysis: the complaint details provided indicate that the stent dislodged off the balloon on insertion through the introducer sheath.The introducer sheath used in this case was not returned.The complaint details also mentioned that there were two devices used in the procedure but were not sure which of the two devices had the stent dislodgement.The returned device balloon surface was evaluated to determine if the stent was crimped properly during manufacturing, when the stent is crimped on to the folded balloon the stent frame leaves impressions of the stent frame on the surface of the balloon.The impressions indicate if the stent was properly crimped on to the balloon.The crimped stent impressions were clearly visible indicating that the stent was properly crimped during manufacturing.The introducer sheath used in the case was not returned.Engineering summary: a full review of the catheter lot history records for the device in question was performed.The records indicate that this lot of catheters passed atrium's final lot qualification testing.This inspection requires that the catheter lot must pass the following: ability of the stent and delivery system to be passed through the labeled introducer sheath.Ability to deploy the stent at nominal pressure (8atm).Ability to withdraw the deflated balloon catheter back through the labeled introducer sheath ability of the delivery system to withstand 10 inflate/deflate cycles at the rated burst pressure (12atm) without leaks or failures.Balloon burst testing.The balloon must burst over the rated burst pressure specified on the label (12atm) manifold to shaft tensile testing.Samples when tensile tested must not break or separate below 3.37lbs result: all (b)(4) quality inspection samples passed this final inspection without any non-conformances noted during the final lot qualification testing related to the stent.All samples passed through the introducer sheath without any stent dislodgments.Conclusion: based on the details of the event and the successful lot qualification test data, atrium can find no fault with the device and or lot of stent delivery systems in question.
 
Event Description
Follow up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATRIUM ICAST COVERED STENT
Type of Device
PROSTHESIS, TRACHEAL, EXPANDABLE
Manufacturer (Section D)
ATRIUM MEDICAL CORPORATION
hudson NH
Manufacturer (Section G)
ATRIUM MEDICAL CORPORATION
5 wentworth drive
hudson NH 03051
Manufacturer Contact
lynda mclaughlin
5 wentworth drive
hudson, NH 03051
6038645470
MDR Report Key4899491
MDR Text Key22243290
Report Number1219977-2015-00181
Device Sequence Number1
Product Code JCT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050814
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/02/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/08/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Model Number85455
Device Catalogue Number85455
Device Lot Number221327
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/23/2015
Was Device Evaluated by Manufacturer? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
7 FR COOK ANSEL SHEATH
Patient Outcome(s) Required Intervention;
Patient Age69 YR
-
-