As reported by physician: in the middle of a colonoscopy, during which i removed a total of nine polyps, we lost all visualization and the screen went stark white.I froze in my position, the risk being that if i can't see, any movement i make could injure the colon.We were able to get the light source working again, however it was much dimmer than usual throughout the rest of the procedure.Additionally, scope buttons lost functionality.I was unable to freeze or capture images or change to fice.We were able to complete the procedure, and there were no untoward complications, gratefully.
|
On 06/10/2015, the subject video processor was received at fujifilm medical systems endoscopy division.Upon inspection, it was observed the unit's video connector has corrosion due to residual fluid, as well as water residue around the top cover.The customer's claims of image loss and loss of endoscope button functionality are consistent with the inspection findings.The introduction of moisture, either by attachment of a wet endoscope connector or on the unit's main cover can damage the unit, potentially leading to image loss and endoscope malfunction.The customer will be re-inserviced on proper handling of endoscopic equipment.
|