The referenced device was returned to olympus medical system corp.(omsc) for evaluation.Omsc evaluated the device and found that the surface glass of the md-631 was broken and the pieces of the glass were dropped into the inside of the light source.It would appear that the damage caused the extinction of the md-631, alternatively the spare lamp was lit, and consequently the endoscopic image became dark.Also, it would appear that the loud noise occurred when the md-631 was broken.Based on the similar cases with the same model, it is known that by inappropriate handling of the md-631 such as the adhesion of the foreign object on the surface glass, strong impact, and/or not secured installation, the md-631 gets broken.The manufacturing history was reviewed, with no irregularities noted.Based upon the finding of the evaluation, this report appears to be related to user handling.There were no further details provided.If significant additional information is received, this report will be supplemented.This report is being submitted as a medical device report in an abundance of caution.
|