Model Number 60-02-60 |
Device Problem
Other (for use when an appropriate device code cannot be identified) (2203)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/11/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
Sorin group (b)(4) manufactures the sorin centrifugal pump 5 (cp5).The incident occurred in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).The investigation is ongoing.A follow-up report will be sent when the investigation is complete.
|
|
Event Description
|
Sorin group received a report that the scp panel switched off.The mains switched off and on several times.Another system was used to complete the case.There was no patient injury.
|
|
Manufacturer Narrative
|
Livanova (b)(4) manufactures the centrifugal pump 5 (cp5).The incident occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).The control panel of the cp5 was returned to livanova (b)(4) for further investigation.Visual inspection did not identify any abnormalities or defects.Functional testing could not reproduce the reported issue.A hardware analysis identified the switch as a potential root cause.The switch connections were disconnected and connected again and then the device was tested for 48 hours without failure.The switch was replaced as a precaution.An exact root cause could not be determined, though a faulty switch is a likely root cause.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.
|
|