Model Number ORCHESTRA PLUS
Device Problems Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Device Displays Incorrect Message (2591)
Patient Problem No Information (3190)
Event Date 07/15/2015
Event Type  malfunction  
Event Description
It was reported that the subject programmer installed with smartview version 2.44 could not be used.A message regarding the initialisation head was displayed permanently.Reboot of the programmer did not resolve the issue.An explanation is required.
 
Event Description
It was reported that the subject programmer installed with smartview version 2.44 could not be used.A message regarding the initialisation head was displayed permanently.Reboot of the programmer did not resolve the issue.An explanation is required.
 
Manufacturer Narrative
Preliminary analysis showed that a sudden shutdown of the programmer had most certainly caused the database corruption leading to the reported behaviour.
 
Event Description
It was reported that the subject programmer installed with smartview version 2.44 could not be used.A message regarding the initialisation head was displayed permanently.Reboot of the programmer did not resolve the issue.An explanation is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORCHESTRA
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
SORIN GROUP ITALIA S.R.L. - CRM FACILITY
parc d'affaires noveos 4 avenue réaumur
.
clamart 92140
FR  92140
Manufacturer (Section G)
KONTRON SUDETENSTR. 8 87600 KAUFBEUREN GERMANY
parc d'affaires noveos 4 avenue réaumur
.
clamart 92140
FR   92140
Manufacturer Contact
elodie vincent
parc d'affaires noveos 4 avenue réaumur
.
clamart 92140
FR   92140
0146013665
MDR Report Key4924674
MDR Text Key6051720
Report Number1000165971-2015-00436
Device Sequence Number1
Product Code DTB
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
PP980049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,health professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/15/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/20/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberORCHESTRA PLUS
Device Catalogue NumberORCHESTRA PLUS
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date07/15/2015
Event Location Hospital
Date Manufacturer Received10/12/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-