Model Number ORCHESTRA PLUS |
Device Problems
Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Device Displays Incorrect Message (2591)
|
Patient Problem
No Information (3190)
|
Event Date 07/15/2015 |
Event Type
malfunction
|
Event Description
|
It was reported that the subject programmer installed with smartview version 2.44 could not be used.A message regarding the initialisation head was displayed permanently.Reboot of the programmer did not resolve the issue.An explanation is required.
|
|
Event Description
|
It was reported that the subject programmer installed with smartview version 2.44 could not be used.A message regarding the initialisation head was displayed permanently.Reboot of the programmer did not resolve the issue.An explanation is required.
|
|
Manufacturer Narrative
|
Preliminary analysis showed that a sudden shutdown of the programmer had most certainly caused the database corruption leading to the reported behaviour.
|
|
Event Description
|
It was reported that the subject programmer installed with smartview version 2.44 could not be used.A message regarding the initialisation head was displayed permanently.Reboot of the programmer did not resolve the issue.An explanation is required.
|
|