MAUDE Adverse Event Report: NAKANISHI, INC. NSK SGA-E2S; HANDPIECE, ROTARY BONE CUTTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NAKANISHI, INC. NSK SGA-E2S; HANDPIECE, ROTARY BONE CUTTING Back to Search Results
Model Number SGA-E2S
Device Problem Overheating of Device (1437)
Patient Problem Burn(s) (1757)
Event Type  malfunction  
Event Description
The following information is from (b)(6) to nakanishi inc.(nsk), regarding a device manufactured by nsk.Event summary: on (b)(6), nsk america received a handpiece (model sga-e2s serial (b)(4)) from dealer for repair.Incoming comments stated "over heating.Burnt patient lip".Immediately (b)(6) contacted dealer for dentist information.(b)(6) contacted office manager at dr.Office, received call back from manager.During conversation it was stated the dentist had burned the patient with the handpiece.No information provided on patient or date of the event.(b)(6) attempted contact by telephone and e-mail on (b)(6) 2015 no response and further follow-ups with no response.(b)(6) replaced and forwarded this handpiece to nhq for full evaluation on 06/17/2015.
 
Manufacturer Narrative
Nakanishi is now investigating this event and will submit additional information including results obtained through the investigation as follow-up report.
 
Manufacturer Narrative
Upon receipt from the dentist of the device involved in the mdr event, nakanishi conducted a failure analysis of the returned device that is included in an attempt to measure the temperature of the operating device.These activities are described in more detail below.Methodology used: as an investigation approach nakanishi inc.(b)(4)(manufacturer) examined the dhr for device (sga-e2s, serial no (b)(4)).There were no problems observed during the manufacturing or testing noted in the dhr; nakanishi conducted a visual inspection of the returned device and performed a simple movement test.There were no visible abnormalities, such as cracks or debris, on the outside of the handpiece.Nakanishi then set a test bur in the handpiece and rotated it by hand.Nakanishi observed that the bur did not rotate smoothly; nakanishi measured the temperature of the handpiece.Nakanishi attached a thermocouple (sensor to measure a temperature) to four testing points.Nakanishi took photographs of the configuration of the temperature measurement test.Nakanishi rotated the handpiece at 40,000 rpm which is maximum rpm for the motor that drives the handpiece without water spray and measured the exothermic situation.The temperature reached 72.7 degrees c, 61.6 degrees c, 33.1 degrees c and 30.5 degrees c respectively, 69 seconds after the beginning of the test.Nakanishi created charts using the data for analysis.Identification of the specific failure mode(s) and/or mechanism(s) and the associated device components involved: nakanishi disassembled the handpiece and performed a visual inspection of the inside parts.Nakanishi observed buildup of shavings from worn ball bearing retainer inside the head cap and the cartridge (head cap side).Nakanishi reassembled and washed the handpiece using nakanishi pana-spray per the instruction included in the user manual.Nakanishi observed dirt/debris expelled from the head of the handpiece by using a white filter to catch anything that was expelled.Nakanishi took photographs of the expelled dirt/debris.Nakanishi believes that overheating was caused by the wear of bearing retainer.The dirt/debris accumulated inside the bearing retainer increased the rear bearing load and caused excess heat generation.Conclusions reached based on the investigation and analysis results: nakanishi identified that the cause of overheating of the returned device was due to damage to the bearing retainer that was caused by accumulated dirt/debris.The damage observed caused abnormal rotational resistance, which would result in the handpiece overheating.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NSK SGA-E2S
Type of Device
HANDPIECE, ROTARY BONE CUTTING
Manufacturer (Section D)
NAKANISHI, INC.
700 shimohinata
kanuma-shi, tochigi-ken 322- 8666
JA  322-8666
Manufacturer (Section G)
NAKANISHI, INC.
700 shimohinata
kanuma-shi, tochigi-ken
JA  
Manufacturer Contact
ken block
1201 richardson dr.
ste 280
richardson, TX 75080
9724809554
MDR Report Key4926323
MDR Text Key6050226
Report Number9611253-2015-00101
Device Sequence Number1
Product Code KMW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K970953
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Dentist
Type of Report Initial
Report Date 07/13/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/17/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberSGA-E2S
Device Catalogue NumberH265
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer06/26/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/26/2015
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/04/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-