Model Number 60-02-50
Device Problem Other (for use when an appropriate device code cannot be identified) (2203)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/21/2015
Event Type  malfunction  
Event Description
Sorin group received a report that the centrifugal pump was not reading flows properly during a procedure.There was no report of patient injury.
 
Manufacturer Narrative
Patient information was not provided.Sorin group (b)(4) manufactures the centrifugal pump system.The incident occurred in (b)(4).This medwatch report is filed on behalf of sorin group (b)(4).Sorin group (b)(4) received a report that the centrifugal pump was not reading flows properly during a procedure.There was no patient injury.The investigation is ongoing.A follow up report will be sent when the investigation is complete.
 
Manufacturer Narrative
Sorin group (b)(4) manufactures the centrifugal pump system.The incident occurred in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).A sorin group field service representative was dispatched to the facility to investigate.The service representative was able to confirm the reported malfunction.Further investigation of the device revealed that the failure was caused by a defective board.The board was replaced to resolve the issue and subsequent testing did not reveal any further failures.The device was returned to service.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.Sorin group (b)(4) will continue to monitor for trends related to this type of issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND GMBH
lindbergh strasse 25
munchen 80939
GM  80939
Manufacturer (Section G)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key4932588
MDR Text Key6015742
Report Number9611109-2015-00276
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K032213
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,health professional
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 06/23/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/17/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-02-50
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received12/09/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/01/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse