(b)(4).One (1) stapler of catalog number 528235 visistat 35w 6/box was received not used, closed in original packaging, lot # 73a1500099 was confirmed.During visual inspection was observed that components looked well assembled, no stuck staples in tip the cartridge, staples was not observed misaligned.Functional inspection: stapler was removed from the original package, after stapler was activated and one staple was loaded, closed & released.Stapler was activated several time and a total of (b)(4) staples were closed properly (no deformed); no quality issues were found during testing.Sample received did not confirm the defect reported by the customer staple deformed during visual inspection.A functional inspection was performed with staples received, the device worked properly.Therefore, a corrective action is not required at this time.In addition, all products are 100% tested by manufacturing and this defect would have been detected during the functional testing.
|