ROCHE DIAGNOSTICS ROCHE CARDIAC D-DIMER; FIBRIN SPLIT PRODUCTS Back to Search Results
Catalog Number 04877802190
Device Problem High Test Results (2457)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/08/2015
Event Type  malfunction  
Event Description
The customer received a questionable d-dimer result for one patient sample from the cobas h232 meter serial number (b)(4).The meter is not sold in the united states.The initial result was 2.40 ug/ml and was reported out.The patient was sent to hospital and approximately two hours later a result of 1.2 ug/ml was received from an unknown analyzer.The patient was not adversely affected.Relevant retention material was tested and the results fulfilled the requirements.
 
Manufacturer Narrative
This event occurred in (b)(6).
 
Manufacturer Narrative
A specific root cause could not be identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROCHE CARDIAC D-DIMER
Type of Device
FIBRIN SPLIT PRODUCTS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 6830 5
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key4936084
MDR Text Key6010872
Report Number1823260-2015-03852
Device Sequence Number1
Product Code GHH
Combination Product (y/n)N
Reporter Country CodeNO
PMA/PMN Number
K033491
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 08/04/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/23/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number04877802190
Device Lot Number28217110
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Event Location Laboratory
Date Manufacturer Received07/08/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-