• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - FREMONT (CE) HYDRO THERMABLATOR (HTA) ENDOMETRIAL ABLATION SYSTEM; DEVICE, THERMAL ABLATION, ENDOMETRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - FREMONT (CE) HYDRO THERMABLATOR (HTA) ENDOMETRIAL ABLATION SYSTEM; DEVICE, THERMAL ABLATION, ENDOMETRIAL Back to Search Results
Model Number M00658001R0
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Patient Involvement (2645)
Event Date 07/01/2015
Event Type  malfunction  
Event Description
It was reported to boston scientific corporation that a genesys hydrothermablation endometrial ablation system was being tested on (b)(6) 2015.According to the complainant, the console failed the biomed electrical safety test and was rendered to be unusable.
 
Manufacturer Narrative
The device has been received for analysis.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Manufacturer Narrative
Analysis of the returned genesys hydrothermablation (hta) endometrial ablation system revealed that the unit passed the leakage & earth resistance safety test and ac current draw test.The console was opened and visually inspected for loose connections and loose or missing hardware.No loose connections or loose or missing hardware were found.As the reported problem was not duplicated during the testing of the genesys console, the most probable root cause is not confirmed.
 
Event Description
It was reported to boston scientific corporation that a genesys hydrothermablation endometrial ablation system was being tested on (b)(6) 2015.According to the complainant, the console failed the biomed electrical safety test and was rendered to be unusable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HYDRO THERMABLATOR (HTA) ENDOMETRIAL ABLATION SYSTEM
Type of Device
DEVICE, THERMAL ABLATION, ENDOMETRIAL
Manufacturer (Section D)
BOSTON SCIENTIFIC - FREMONT (CE)
47215 lakeview blvd
north dock
fremont CA 94538
Manufacturer (Section G)
BOSTON SCIENTIFIC - FREMONT (CE)
47215 lakeview blvd
north dock
fremont CA 94538
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key4939792
MDR Text Key17220175
Report Number3005099803-2015-02100
Device Sequence Number1
Product Code MNB
Combination Product (y/n)N
PMA/PMN Number
P000040
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative,company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 07/01/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/24/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberM00658001R0
Device Catalogue Number58001R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/10/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received09/03/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/05/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-