Catalog Number UNK_SEL
Device Problems Bent (1059); Loose or Intermittent Connection (1371)
Patient Problems Swelling (2091); Injury (2348)
Event Date 07/01/2015
Event Type  Injury  
Event Description
The customer reported that the patient had experienced a serious clavicle injury.A lock down band and a stryker lateral clavicle plate was used to assist in healing the injury.It was reported that due to the position of the clavicle plate, the screws are getting loose and are moving, causing a subcutaneous swelling.Moreover, towards the ac joint, the clavicle bends a bit downwards.The customer alleges that these issues would not arise if the distal part of the lateral clavicle plate would be bent a bit downwards.
 
Manufacturer Narrative
Device will not be returned.If additional information becomes available it will be provided on a supplemental report.
 
Manufacturer Narrative
Even though the device was not returned for inspection, the reported incident of getting loose screws (s-41 - poor fixation) could be confirmed thanks to the x-rays provided by the customer.Based on the currently available information, the root cause can be attributed to a user-related issue.Our clinical expert was involved in this investigation and reported the following in his clinical statement: "all x-rays show unstable lateral clavicle fractures involving the ac joint.In all three cases the lateral fragment with articulation to the ac joint is very small.Such small lateral fragments are a borderline indication for plate fixation in general, including the variax superior lateral clavicle plates because the bone quality is poor (even in younger sportive patients) and the loading on the screws very close to the ac joint is very high.For such ¿very lateral¿ fractures a variax hook plate would be a much better alternative solution.When a superior lateral clavicle plate is used for treatment of such fractures, the lateral locking screws should be inserted with divergent axes to prevent back out via form fit.In the available x-rays the axes of the backed out screws look nearly parallel to each other, but only one x-ray view is presented.Nevertheless, even with divergent placement of the lateral locking screws postoperative mobilization has to be done with great care.Such orif does not allow for early active rehabilitation in the first 6 weeks until significant callus formation is visible on the follow up x-rays.Furthermore, it has been reported that 'the customer alleges that these issues would not arise if the distal part of the lateral clavicula plate would be bend a bit downwards'.All variax plates are anatomically pre-shaped plates, which match to an ¿average anatomy¿.For adaption of the plate to the individual anatomy there are bending irons in the variax instrument tray, which allow for easy shaping of the plate to achieve optimal fit to the bone." a review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.Indications for any material, manufacturing or design related problems were not determined as the lot number was not communicated.If the device is returned for investigation or if further information is available, the investigation will be reassessed.
 
Event Description
The customer reported that the patient had experienced a serious clavicle injury.A lock down band and a stryker lateral clavicle plate was used to assist in healing the injury.It was reported that due to the position of the clavicle plate, the screws are getting loose and are moving, causing a subcutaneous swelling.Moreover, towards the ac joint, the clavicle bends a bit downwards.The customer alleges that these issues would not arise if the distal part of the lateral clavicle plate would be bent a bit downwards.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN_SELZACH_PRODUCT
Type of Device
IMPLANT
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
CH  2545
Manufacturer (Section G)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
CH   2545
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key4939942
MDR Text Key6036089
Report Number0008031020-2015-00327
Device Sequence Number1
Product Code HRS
Combination Product (y/n)N
Reporter Country CodeNL
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,health professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/01/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/24/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberUNK_SEL
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received10/22/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-