• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC. FREESTYLE; BLOOD GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC. FREESTYLE; BLOOD GLUCOSE MONITORING SYSTEM Back to Search Results
Catalog Number 11001-1
Device Problem False Reading From Device Non-Compliance (1228)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/02/2015
Event Type  malfunction  
Manufacturer Narrative
The product has been requested back for investigation.A follow- up report will be submitted once additional information is obtained.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
Customer reported receiving erratic readings on their omni pod system while using freestyle test strips.Customer reported receiving readings of 373 mg/dl, 41 mg/dl and 393 mg/dl within 10 minutes.The results when plotted on a parkes error grid fell into the "d" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
 
Manufacturer Narrative
The reported test strips were returned and investigated with retained meter.The complaint is not confirmed and no new issues were observed.All results were within the range specification and no errors were observed during control solution testing.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE
Type of Device
BLOOD GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC.
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
meg manager
1360 south loop road
alameda, CA 94502-7001
5108644391
MDR Report Key4940779
MDR Text Key22889223
Report Number2954323-2015-00365
Device Sequence Number1
Product Code LFR
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K092638
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Report Date 08/20/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/24/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date08/31/2016
Device Catalogue Number11001-1
Device Lot Number1501502
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/21/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received08/04/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/15/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-