The minuteman ups was not returned to the manufacturer for evaluation.It was reported that the facility biomedical engineer opened the minuteman ups and observed accumulated lint which is likely the material that ignited.The specification of the temporary and the replacement ups installed by the facility biomedical engineer was not provided to the manufacturer.It was a non microscan ups and not supported by beckman coulter.A microscan minuteman ups specification was provided to the customer.The manufacturer customer technical support explained that ups with less than 2000w would not support the walkaway instrument.At the time of the event, it was reported that the customer was loading 4-panels into the walkaway.The manufacturer customer technical support (cts) recommended connecting the instrument to the emergency line however the customer declined.The customer agreed to the cts recommendation to set up the 4-panels that were currently being loaded, add oil manually and incubate offline in an external incubator at 35°c for 18 hours, dispense the reagents next day and perform manual read.No further action is required.(b)(4).
|