U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENOVATE MEDICAL MOBIUS POWER MEDICAL CART BATTERY; LITHIUM-ION BATTERY FOR CART

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENOVATE MEDICAL MOBIUS POWER MEDICAL CART BATTERY; LITHIUM-ION BATTERY FOR CART Back to Search Results
Device Problems Fire (1245); Smoking (1585); Noise, Audible (3273)
Patient Problem No Code Available (3191)
Event Date 06/04/2015
Event Type  Injury  
Event Description
At 0144 in the morning, a lithium-ion battery used to power a hospital workstation on wheels caught fire while in its wall mounted charging station.A hospital staff member heard hissing and popping, followed by smoke then fire; 911 was called and the hospital's intensive care pts had to be evacuated.Five hospital staff members reported inhaling smoke, but no injuries otherwise resulted.Clean-up on the unit took a week to complete.This is the second time this year a battery from the same company has resulted in a fire within our hospital system.While the battery is not a medical device per se, the malfunction affected pt care.An ecri alert was released related to the matter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOBIUS POWER MEDICAL CART BATTERY
Type of Device
LITHIUM-ION BATTERY FOR CART
Manufacturer (Section D)
ENOVATE MEDICAL
murfreesboro TN
MDR Report Key4945812
MDR Text Key6256282
Report NumberMW5044781
Device Sequence Number1
Product Code FCO
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/21/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/21/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/05/2015
Is the Reporter a Health Professional? Yes
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
-
-