Concomitant: product id 64002, lot# n342445, implanted: 2014-(b)(6), product type adapter.Product id 782a51, serial# (b)(4), implanted: 2011-(b)(6), product type extension.Product id 7482a51, serial# (b)(4), implanted: 2011-(b)(6), product type extension.Product id 3387s-40, lot# v621406, implanted: 2011-(b)(6), product type lead.(b)(4).
|
The patient experienced considerable improvements with the treatment of the device.The end of service (eos) code displayed while usual telemetry with the clinician programmer was done.The same showed a message saying that the battery of the implantable neurostimulator (ins) was run out and that it has to be changed.Premature battery depletion of the ins occurred.The ins had switched off and was tried to switch it on, but it was not possible.The patient had started shivering again.The ins lasted eight months from the implant date to the end of life (eol).Impedance measuring of the poles to prove the state of the electrodes and its connections, showed normal/correct values.It was further reported that a battery longevity estimate was performed using the following settings taken from the clinician programmer: l amplitude: 4.2v pulsewidth: 180's rate: 170hz therapy impedance: 782ohms r amplitude: 4.2v pulsewidth: 180rate: 170hz therapy i mpedance: 811ohms the approximate longevity from the beginning of life to elective replacement indicator (eri) was 8.21 months.The approximate longevity from the beginning of life to eos was 11.21 months.On the clinician programmer report the battery voltage was 2.94v which indicated that there was a significant voltage bounce back after the battery reached eos and turned off.The outcome remains unknown at this time.If additional information is received, a follow up report will be sent.
|